| Normal operating
settings are - | 
      50kV; 30mA. | 
    
| Typical stand-by
settings are - | 
      40kV; 10mA. (short term, i.e. a few hours). | 
| 20kV; 5mA (longer term). | 
![]()  | 
    
| Nonius FR-590 sealed-tube x-ray generator control panel | 
  If the display shows 30mA, 50kV and 1.5kW, then
  the power is
  already up.
  
  If the display shows less than this then you need
  to crank
  up the power as follows:
  
  Press the kV button
  (middle right
  above), then enter "50" on the keypad at the right of the panel and press the "ENTER" key.
  Press the mA button
  (middle left
  above), then enter "30" on the keypad at the right of the panel and press the "ENTER" key.
  
  If the x-rays are off, then push the "ON" button on the left of the
  panel. 
  The x-rays will come on to low power (20kV; 5mA). 
  From
  here you can crank the power to 40kV; 10mA and then on
  to 50kV, 30mA as described above. 
  
  If the generator was off, you
  need to
  turn the key to the horizontal position before turning the x-rays on
  and
  cranking the power up.
  