Normal operating
settings are - |
50kV; 30mA. |
Typical stand-by
settings are - |
40kV; 10mA. (short term, i.e. a few hours). |
20kV; 5mA (longer term). |
Nonius FR-590 sealed-tube x-ray generator control panel |
If the display shows 30mA, 50kV and 1.5kW, then
the power is
already up.
If the display shows less than this then you need
to crank
up the power as follows:
Press the kV button
(middle right
above), then enter "50" on the keypad at the right of the panel and press the "ENTER" key.
Press the mA button
(middle left
above), then enter "30" on the keypad at the right of the panel and press the "ENTER" key.
If the x-rays are off, then push the "ON" button on the left of the
panel.
The x-rays will come on to low power (20kV; 5mA).
From
here you can crank the power to 40kV; 10mA and then on
to 50kV, 30mA as described above.
If the generator was off, you
need to
turn the key to the horizontal position before turning the x-rays on
and
cranking the power up.